1.

Background

Due to the high prevalence of hemolysis in specimens received from the emergency department (ED), several strategies have been proposed to improve sample quality, but none of these seem effective to overcome the problem. In a preliminary study we showed that the use of S-Monovette blood collection system was effective to lower the risk of hemolysis in venous blood samples collected from intravenous catheters. This study was hence aimed to verify whether the replacement of a conventional vacuum system with S-Monovette may be effective to reduce the burden of hemolysis in the daily practice of a large urban ED.

Materials and methods

The study was divided in two observational periods of 4 months each. In the former period, blood was collected from intravenous catheters using BD Vacutainer SST II Plus plastic serum tubes, whereas in the latter period the blood was drawn from intravenous catheters using S-Monovette blood tubes in aspiration mode. Sample hemolysis was automatically assessed in all serum samples by photometrical measurement.

Results

The total number of hemolysed serum specimens was 624/14155 (4.41%) in the first phase of the study, and 342/13319 (2.57%) in the second phase of the study (P < 0.001).

Conclusion

Results of our study confirm that the introduction of the Sarstedt S-Monovette blood tubes has reduced the hemolysis rate in the emergency department compared to the previously used BD Vacutainer® SST II Plus plastic serum tubes.Key words: preanalytical phase, hemolysis, blood specimen collection, catheter, emergency department  相似文献   

2.

Background:

In vitro hemolysis can be induced by several biological and technical sources, and may be worsened by forced aspiration of blood in vacuum tubes. This study was aimed to compare the probability of hemolysis by drawing blood with a commercial evacuated blood collection tube, and S-Monovette used either in the “vacuum” or “aspiration” mode.

Materials and methods:

The study population consisted in 20 healthy volunteers. A sample was drawn into 4.0 mL BD Vacutainer serum tube from a vein of one upper arm. Two other samples were drawn with a second venipuncture from a vein of the opposite arm, into 4.0 mL S-Monovette serum tubes, by both vacuum an aspiration modes. After separation, serum potassium, lactate dehydrogenase (LD) and hemolysis index (HI) were tested on Beckman Coulter DxC.

Results:

In no case the HI exceed the limit of significant hemolysis. As compared with BD Vacutainer, no significant differences were observed for potassium and LD using S-Monovette with vacuum method. Significant increased values of both parameters were however found in serum collected into BD Vacutainer and S-Monovette by vacuum mode, compared to serum drawn by S-Monovette in aspiration mode. The mean potassium bias was 2.2% versus BD Vacutainer and 2.4% versus S-Monovette in vacuum mode, that of LD was 2.7% versus BD Vacutainer and 2.1% versus S-Monovette in vacuum mode. None of these variations exceeded the allowable total error.

Conclusions:

Although no significant macro-hemolysis was observed with any collection system, the less chance of producing micro-hemolysis by S-Monovette in aspiration mode suggest that this device may be used when a difficult venipuncture combined with the vacuum may increase the probability of spurious hemolysis.  相似文献   

3.

Introduction:

Blood collection through intravenous lines frequently causes spurious hemolysis. Due to specific structure, the tube holder Holdex® (Greiner Bio-One GmbH, Kremsmuenster, Austria) is supposed to prevent erythrocyte injury in samples collected from catheters, so that we planned a specific study to support this hypothesis.

Materials and methods:

Blood was collected from emergency department (ED) patients with 20-gauge catheter. In patients with odd order numbers, first and second tubes were collected with conventional holder (BD Vacutainer One Use Holder, Becton Dickinson, Milan, Italy) and the third with Holdex, whereas in even patients first and second tubes were drawn with Holdex and the third using BD Vacutainer One Use Holder. The first tube was discarded, whereas the second and third were centrifuged and serum was tested for potassium, lactate dehydrogenase (LD) and hemolysis index.

Results:

The final study population consisted in 60 ED patients. Concentrations of potassium (4.25 vs. 4.16 mmol/L; P = 0.031), LD (498 vs. 459 U/L; P = 0.039) and cell-free hemoglobin (0.42 vs. 0.22 g/L; P = 0.042) were higher in samples collected with BD Vacutainer One Use Holder than with Holdex. The mean bias of cell-free hemoglobin was −0.4 g/L in samples collected with Holdex. Although the frequency of samples with cell-free hemoglobin > 0.5 g/L was identical (17/60 vs. 17/60; P = 1.00), the frequency of those with concentrations >3.0 g/L was higher using BD Vacutainer One Use Holder than Holdex (4/60 vs. 0/60; P = 0.042).

Conclusions:

The use of Holdex for drawing blood from intravenous lines may be effective for reducing gross hemolysis.  相似文献   

4.

Introduction

This study aimed to assess confidence level of healthcare professionals in venipuncture and their knowledge on the possible causes of in vitro hemolysis.

Materials and methods

A sample of 94 healthcare professionals (nurses and laboratory technicians) participated in this survey study. A four-section questionnaire was used as a research instrument comprising general information for research participants, knowledge on possible causes of in vitro hemolysis due to type of material used and venipuncture technique and specimen handling, as well as assessment of healthcare professionals’ confidence level in their own ability to perform first and last venipuncture.

Results

The average score on the knowledge test was higher in nurses’ than in laboratory technicians (8.11 ± 1.7, and 7.4 ± 1.5, respectively). The difference in average scores was statistically significant (P = 0.035) and Cohen’s d in the range of 0.4 indicates that there is a moderate difference on the knowledge test among the health care workers. Only 11/94 of healthcare professionals recognized that blood sample collection from cannula and evacuated tube is method which contributes most to the occurrence of in vitro hemolysis, whereas most risk factors affecting occurrence of in vitro hemolysis during venipuncture were recognized. There were no significant differences in mean score on the knowledge test in relation to the confidence level in venipuncture (P = 0.551).

Conclusion

Confidence level at last venipuncture among both profiles of healthcare staff was very high, but they showed insufficient knowledge about possible factors affecting hemolysis due to materials used in venipuncture compared with factors due to venipuncture technique and handling of blood sample.Key words: hemolysis, patient safety, phlebotomy, medical staff, questionnaire, preanalytical phase  相似文献   

5.

Background

It is still uncertain whether or not avoidance to let disinfectant alcohol dry at the site of venipuncture is a source of spurious hemolysis when drawing venous blood.

Methods:

In a consecutive series of 52 outpatients referred for routine laboratory testing, venous blood was drawn by direct venipuncture with (odd group) or without (pair group) wiping 70% isopropyl alcohol at the site of venipuncture. A 3.5 mL evacuated tube with clot activator and gel separator was drawn from a vein of the upper limb, serum was immediately separated with standard centrifugation and tested for potassium, lactate dehydrogenase (LD), aspartate aminotransferase (AST) and hemolysis index (HI) on Roche Cobas.

Results:

No specimen was discarded for unsatisfactory venipuncture. No differences for age and gender were observed between groups. As regards the four parameters investigated, no significant differences could be observed between patients in whom blood was drawn with or without letting the alcohol dry. It is also noteworthy that no sample in both groups exceeded the conventional sample rejection threshold of cell-free hemoglobin.

Conclusions:

The results of our prospective, randomized study attest that failure to wipe alcohol at the site of venipuncture should not be considered as a potential source of spurious hemolysis when drawing blood.  相似文献   

6.

Introduction:

Sample type recommended by the manufacturer for the digoxin Abbott assay is either serum collected in glass tubes or plasma (sodium heparin, lithium heparin, citrate, EDTA or oxalate as anticoagulant) collected in plastic tubes. In our hospital samples are collected in plastic tubes. Our hypothesis was that the serum sample collected in plastic serum tube can be used interchangeably with plasma sample for measurement of digoxin concentration. Our aim was verification of plastic serum tubes for determination of digoxin concentration.

Materials and methods:

Concentration of digoxin was determined simultaneously in 26 venous blood plasma (plastic Vacuette, LH Lithium heparin) and serum (plastic Vacuette, Z Serum Clot activator; both Greiner Bio-One GmbH, Kremsmünster, Austria) samples, on Abbott AxSYM analyzer using the original Abbott Digoxin III assay (Abbott, Wiesbaden, Germany). Tube comparability was assessed using the Passing Bablok regression and Bland-Altman plot.

Results:

Serum and plasma digoxin concentrations are comparable. Passing Bablok intercept (0.08 [95% CI = −0.10 to 0.20]) and slope (0.99 [95% CI = 0.92 to 1.11]) showed there is no constant or proportional error.

Conclusion:

Blood samples drawn in plastic serum tubes and plastic plasma tubes can be interchangeably used for determination of digoxin concentration.  相似文献   

7.

Introduction:

Serum blood collection tubes with separator gel are widely used by many laboratories for chemistry analyses. We describe a case of a primary blood collection tube filled with blood sample and a floating separator gel.

Materials and methods:

The blood sample was collected from a 51 years old female in intensive care unit with the diagnosis of pneumonia into a BD Vacutainer SST tube (Becton Dickinson, NJ, USA) containing serum separator gel and conveyed to the core laboratory of Marmara University Hospital within 30 minutes from collection. Sample was immediately centrifuged at room temperature at 1500 × g for 10 minutes.

Results:

The analyses revealed a highly increased total protein concentration of 145 g/L (reference interval 64–83 g/L). The nephelometric analyses showed an elevated serum IgG concentration of 108 g/L (reference interval 6.5–16 g/L) and IgG lambda monoclonal band was determined by serum immunofixation electrophoresis.

Conclusion:

Limitation of the separator gel tubes in patients with a high plasma density and its possible effects on test results and laboratory costs should be remembered. The clinical diagnosis stated in the information system should also reveal known comorbid conditions besides the apparent admission reason. This information would avoid resampling, additional testing, and communication efforts with the clinicians.  相似文献   

8.

Introduction:

Non insulin dependent diabetes mellitus is the most common type of diabetes. Genetic factors, lipid profiles, hypertension are potential risk factors for diabetes mellitus. Adenosine binding cassette transporter proteins 1 (ABCA1) plays a role in cholesterol metabolism, especially high density lipoprotein (HDL-cholesterol). There are multiple mechanisms by which HDL-cholesterol can be atheroprotective, it is clear that the relative activity of ABCA1 plays a major role. We aimed to investigate association of ABCA1 C69T gene polymorphism with lipid levels in Turkish type 2 diabetic patients.

Materials and methods:

After isolation of DNA by ethanol precipitation we determined ABCA1 gene polymorphism by using polimerase chain reaction - restriction fragment lenght polymorphism (PCR-RFLP) method in 107 type 2 diabetic patients and 50 healthy controls.

Results:

We have observed that the frequency of TT genotype is significantly higher in healthy controls compared to patients (14% vs. 3%; P = 0.008). Also frequency of T allele was higher in controls than in patients (34% vs. 21%; P = 0.020; OR (95% CI) = 0.52 (0.30–0.88)). There was no association of lipid levels and ABCA1 C69T polymorphism subgroups.

Conclusion:

We have found significantly higher frequency of both T allele and genotype in control group when compared to patients that made us think that T allele may be a protective factor against diabetes mellitus. But, we could not find a relationship between genotypes and lipid concentrations in our two groups. Larger studies will help us to understand the relationship between ABCA1 C69T genotype and lipid parameters in diabetes mellitus.  相似文献   

9.

Introduction

Although prostatitis is a common male urinary tract infection, clinical diagnosis of prostatitis is difficult. The developmental mechanism of prostatitis is not yet unraveled which led to the elaboration of various biomarkers. As changes in asparagine-linked-(N-)-glycosylation were observed between healthy volunteers (HV), patients with benign prostate hyperplasia and prostate cancer patients, a difference could exist in biochemical parameters and urinary N-glycosylation between HV and prostatitis patients. We therefore investigated if prostatic protein glycosylation could improve the diagnosis of prostatitis.

Materials and methods

Differences in serum and urine biochemical markers and in total urine N-glycosylation profile of prostatic proteins were determined between HV (N = 66) and prostatitis patients (N = 36). Additionally, diagnostic accuracy of significant biochemical markers and changes in N-glycosylation was assessed.

Results

Urinary white blood cell (WBC) count enabled discrimination of HV from prostatitis patients (P < 0.001). Urinary bacteria count allowed for discriminating prostatitis patients from HV (P < 0.001). Total amount of biantennary structures (urinary 2A/MA marker) was significantly lower in prostatitis patients compared to HV (P < 0.001). Combining the urinary 2A/MA marker and urinary WBC count resulted in an AUC of 0.79, 95% confidence interval (CI) = (0.70–0.89) which was significantly better than urinary WBC count (AUC = 0.70, 95% CI = [0.59–0.82], P = 0.042) as isolated test.

Conclusions

We have demonstrated the diagnostic value of urinary N-glycosylation profiling, which shows great potential as biomarker for prostatitis. Further research is required to unravel the developmental course of prostatic inflammation.Key words: diagnostic marker, prostatitis, urinary asparagine-linked glycosylation  相似文献   

10.

Background:

Presently the necessity of fasting time for coagulation tests is not standardized. Our hypothesis is that this can harm patient safety. This study is aimed at evaluating whether a light meal (i.e. breakfast) can jeopardize laboratory coagulation tests.

Materials and methods:

A blood sample was firstly collected from 17 fasting volunteers (12 h). Immediately after blood collection, the volunteers consumed a light meal. Then samples were collected at 1, 2 and 4 h after the meal. Coagulation tests included: activated partial thromboplastin time (APTT), prothrombin time (PT), fibrinogen (Fbg), antithrombin III (AT), protein C (PC) and protein S (PS). Differences between samples were assessed by Wilcoxon ranked-pairs test. The level of statistical significance was set at P < 0.05. Mean % differences were determined and differences between and baseline and 1, 2 and 4h samples were compared with reference change value (RCV).

Results:

A significantly higher % activity of AT was observed at 1 h and 4 h after meal vs. baseline specimen [113 (104–117) and 111 (107–120) vs. 109 (102–118), respectively; P = 0.029 and P = 0.016]. APTT at 2 h was found significantly lower than baseline samples [32.0 (29.9–34.8) vs. 34.1 (32.2–35.2), respectively; P = 0.041]. The results of both Fbg and PS tests were not influenced by a light meal. Furthermore, no coagulation tests had significant variation after comparison with RCV.

Conclusion:

A light meal does not influence the laboratory coagulation tests we assessed, but we suggest that the laboratory quality managers standardize the fasting time for all blood tests at 12 hours, to completely metabolize the lipids intake.  相似文献   

11.

Background

The preanalytical phase represents the major source of variability in laboratory diagnostics. Our aim was to assess to what extent underfilling of primary blood tubes may impact upon routine coagulation testing.

Materials and methods:

Blood was drawn by syringe from 21 healthy volunteers and 6 patients on warfarin therapy, and immediately transferred into 3.6 mL vacuum tubes containing 3.2% sodium citrate (Terumo Europe N.V., Leuven, Belgium). All tubes were filled using standardized volumes of whole blood to produce scalar amounts of filling: 3.6 mL (i.e., 100%), 3.2 mL (89%), 2.8 mL (78%) and 2.4 mL (67%). Samples were mixed and centrifuged at 1300 × g for 10 min. The plasma was tested for prothrombin time (PT), activated partial thromboplastin time (APTT) and fibrinogen (FBG) on ACL TOP (Instrumentation Laboratory - IL, Milan, Italy), using IL reagents. A polynomial plot was derived for each parameter from interpolation of clotting values obtained with different percentages of filling, and these plots were compared with quality specifications (± 2.0 for PT, ± 2.3 for APTT and ± 4.8 for FBG) to calculate the minimal filling volume required to produce clinically acceptable results.

Results:

The equations were (PF, Percentage of filling): PT (sec) = 3.375 × PF^2–6.255 × PF + 17.806 (r = 0.980); APTT (sec) = 8.925 × PF^2–23.578 × PF + 46.356 (r = 0.979); and FBG (mg/dL) = −311.5 × PF^2 + 422.1 × PF + 147.07 (r = 0.994). According to these equations, the minimum allowed thresholds of blood tubes filling were > 61% for PT, > 87% for APTT and > 71% for FBG.

Conclusions:

Our results confirm that routine coagulation testing performed on underfilled tubes may generate biased and clinically misleading test results. This is particularly critical for APTT, wherein tubes filled at less than ∼90% generate unreliable data. The FBG and the PT seem more resistant to underfilling, clinical significant biases being observed only where blood tubes were filled at less than ∼60 and ∼70%, respectively.  相似文献   

12.

Introduction:

Preanalytical errors, along the process from the beginning of test requests to the admissions of the specimens to the laboratory, cause the rejection of samples. The aim of this study was to better explain the reasons of rejected samples, regarding to their rates in certain test groups in our laboratory.

Materials and methods:

This preliminary study was designed on the rejected samples in one-year period, based on the rates and types of inappropriateness. Test requests and blood samples of clinical chemistry, immunoassay, hematology, glycated hemoglobin, coagulation and erythrocyte sedimentation rate test units were evaluated. Types of inappropriateness were evaluated as follows: improperly labelled samples, hemolysed, clotted specimen, insufficient volume of specimen and total request errors.

Results:

A total of 5,183,582 test requests from 1,035,743 blood collection tubes were considered. The total rejection rate was 0.65 %. The rejection rate of coagulation group was significantly higher (2.28%) than the other test groups (P < 0.001) including insufficient volume of specimen error rate as 1.38%. Rejection rates of hemolysis, clotted specimen and insufficient volume of sample error were found to be 8%, 24% and 34%, respectively. Total request errors, particularly, for unintelligible requests were 32% of the total for inpatients.

Conclusions:

The errors were especially attributable to unintelligible requests of inappropriate test requests, improperly labelled samples for inpatients and blood drawing errors especially due to insufficient volume of specimens in a coagulation test group. Further studies should be performed after corrective and preventive actions to detect a possible decrease in rejecting samples.  相似文献   

13.

Introduction

Total 25-hydroxyvitamin D [25(OH)D] is the most reliable indicator of vitamin D status. In this study, we compared two automated immunoassay methods, the Abbott Architect 25-OH Vitamin D assay and the Roche Cobas Vitamin D total assay, with the liquid chromatography-tandem mass spectrometry (LC-MS/MS).

Materials and methods

One hundred venous blood samples were randomly selected from routine vitamin D tests. Two of the serum aliquots were analyzed at the Abbott Architect i2000 and the Roche Cobas 6000’s module e601 in our laboratory within the same day. The other serum aliquots were analyzed at the LC-MS/MS in different laboratory. Passing-Bablok regression analysis and Bland-Altman plot were used to compare methods. Inter-rater agreement was analyzed using kappa (κ) analysis.

Results

The Roche assay showed acceptable agreement with the LC-MS/MS based on Passing-Bablok analysis (intercept: -5.23 nmol/L, 95% CI: -8.73 to 0.19; slope: 0.97, 95% CI: 0.77 to 1.15). The Abbott assay showed proportional (slope: 0.77, 95% CI: 0.67 to 0.85) and constant differences (intercept: 17.08 nmol/L; 95% CI: 12.98 to 21.39). A mean bias of 15.1% was observed for the Abbott and a mean bias of -14.1% was observed for the Roche based on the Bland-Altman plots. We found strong to nearly perfect agreement in vitamin D status between the immunoassays and LC-MS/MS. (κ: 0.83 for Abbott, κ: 0.93 for Roche) using kappa analysis.

Conclusion

Both immunoassays demonstrated acceptable performance, but the Roche Cobas assay demonstrated better performance than the Abbott Architect in the studied samples.Key words: 25-Hydroxyvitamin D, chromatography, immunoassay, methods, tandem mass spectrometry  相似文献   

14.

Introduction

Urinary tract infection (UTI) is one of the most common types of infection. Currently, diagnosis is primarily based on microbiologic culture, which is time- and labor-consuming. The aim of this study was to assess the diagnostic accuracy of urinalysis results from UriSed (77 Electronica, Budapest, Hungary), an automated microscopic image-based sediment analyzer, in predicting positive urine cultures.

Materials and methods:

We examined a total of 384 urine specimens from hospitalized patients and outpatients attending our hospital on the same day for urinalysis, dipstick tests and semi-quantitative urine culture. The urinalysis results were compared with those of conventional semi-quantitative urine culture.

Results:

Of 384 urinary specimens, 68 were positive for bacteriuria by culture, and were thus considered true positives. Comparison of these results with those obtained from the UriSed analyzer indicated that the analyzer had a specificity of 91.1%, a sensitivity of 47.0%, a positive predictive value (PPV) of 53.3% (95% confidence interval (CI) = 40.8–65.3), and a negative predictive value (NPV) of 88.8% (95% Cl = 85.0–91.8%). The accuracy was 83.3% when the urine leukocyte parameter was used, 76.8% when bacteriuria analysis of urinary sediment was used, and 85.1% when the bacteriuria and leukocyturia parameters were combined. The presence of nitrite was the best indicator of culture positivity (99.3% specificity) but had a negative likelihood ratio of 0.7, indicating that it was not a reliable clinical test.

Conclusions:

Although the specificity of the UriSed analyzer was within acceptable limits, the sensitivity value was low. Thus, UriSed urinalysis results do not accurately predict the outcome of culture.  相似文献   

15.

Introduction

Glycolysis affects glucose determination in vitro. The placement of sample tubes in ice-water slurry with plasma separation within 30 minutes is recommended, or alternatively the use of a glycolysis inhibitor. The aim of our two-steps study was to evaluate which Terumo tube is best for glucose determination in routine clinical setting.

Materials and methods

In the first study, blood from 100 volunteers was collected into lithium heparin (LH), NaF/Na heparin (FH) and NaF/citrate buffer/Na2EDTA (FC-Mixture) tubes. LH sample was treated as recommended and considered as reference, while FH and FC-Mixture samples were aliquoted, maintained at room temperature (RT) for 1, 2 and 4 hours; centrifuged and plasma analysed in triplicate. In the second study, samples from 375 volunteers were collected in LH, FH and FC-Mixture tubes and held at RT before centrifugation from 10 to 340 minutes, depending on each laboratory practice. Samples were analysed in one analytical run.

Results

In the first study, FH glucose concentrations were 5.15 ± 0.66 mmol/L, 5.05 ± 0.65 mmol/L and 5.00 ± 0.65 mmol/L (P < 0.001) in tubes stored at RT for 1, 2 and 4 hours, respectively. Mean biases in all time points exceeded the analytical goal for desirable bias based on biological variation criteria. FC-Mixture glucose concentrations were 5.48 ± 0.65 mmol/L, 5.46 ± 0.6 mmol/L and 5.46 ± 0.64 mmol/L in tubes stored at RT for 1, 2 and 4 hours, respectively. Mean biases for FC-Mixture glucose in all time points reached optimal analytical goals. In the second study, the biases for LH and FH glucose compared to reference FC-Mixture glucose exceeded the preset analytical goals, regardless of the blood collection to centrifugation time interval.

Conclusions

FC-mixture tubes glucose concentrations were preserved up to 4h storage at RT. We confirmed that NaF alone does not allow immediate glycolysis inhibition in real life pre-centrifugation storage conditions (up to 340 minutes). FC-Mixture should be used exclusively for glucose determination in laboratories unable to implement the recommended blood samples’ treatment.Key words: glucose, pre-analytical phase, sodium fluoride, citrate acidification, stability  相似文献   

16.

Introduction:

The activities involving phlebotomy, a critical task for obtaining diagnostic blood samples, are poorly studied as regards the major sources of errors and the procedures related to laboratory quality control. The aim of this study was to verify the compliance with CLSI documents of clinical laboratories from South America and to assess whether teaching phlebotomists to follow the exact procedure for blood collection by venipuncture from CLSI/NCCLS H03-A6 - Procedures for the Collection of Diagnostic Blood Specimens by Venipuncture might improve the quality of the process.

Materials and methods:

A survey was sent by mail to 3674 laboratories from South America to verify the use of CLSI documents. Thirty skilled phlebotomists were trained with the CLSI H03-A6 document to perform venipuncture procedures for a period of 20 consecutive working days. The overall performances of the phlebotomists were further compared before and after the training program.

Results:

2622 from 2781 laboratories that did answer our survey used CLSI documents to standardize their procedures and process. The phlebotomists’ training for 20 days before our evaluation completely eliminated non-conformity procedures for: i) incorrect friction of the forearm, during the cleaning of the venipuncture site to ease vein location; ii) incorrect sequence of vacuum tubes collection; and iii) inadequate mixing of the blood in primary vacuum tubes containing anticoagulants or clot activators. Unfortunately the CLSI H03-A6 document does not caution against both unsuitable tourniquet application time (i.e., for more than one minute) and inappropriate request to clench the fist repeatedly. These inadequate procedures were observed for all phlebotomists.

Conclusion:

We showed that strict observance of the CLSI H03-A6 document can remarkably improve quality, although the various steps for collecting diagnostic blood specimens are not a gold standard, since they may still permit errors. Tourniquet application time and forearm clench should be verified by all quality laboratory managers in the services. Moreover, the procedure for collecting blood specimens should be revised to eliminate this source of laboratory variability and safeguard the quality.  相似文献   

17.

Introduction

The collected and shipped blood samples are exposed to a various extra-analytical factors prior to analysis. The aim of the study was to determine the stability of analytes in serum gel tubes and plain tubes exposed to a range of storage temperatures and times after centrifugation.

Materials and methods:

Fifteen healthy volunteers were recruited and venous blood was collected into four tubes, two with and two without gel separator. Analyzing the baseline samples in 30 min, all were stored at 4ºC or 24ºC for 6, 12, 18, 24, 30, 36, 48 and 72 hours and 1 week. Sixteen biochemical anaytes were measured on each sample. Variations remained under the desirable bias considered as clinically insignificant.

Results:

On day three, most analytes remained stable including albumin, protein, creatinine, cholesterol, triglycerides, gamma-glutamyl transferase (GGT), alkaline phosphatase (ALP), alanine aminotransferase (ALT), creatine kinase (CK), lactate dehydrogenase (LD) regardless of tube types. Glucose concentration decreased markedly (P = 0.001) beginning from the first hours of storage in plain serum. The stability maximized for the analytes including glucose, total bilirubin, urea nitrogen (BUN), uric acid stored at 4 ºC in gel tubes. Aspartate aminotransferase (AST) activity increased significantly (P = 0.002) up to 48-h, however bias was not significant clinically. High density lipoprotein (HDL) concentration was stable in gel tubes at 24 ºC, in plain tubes at 4 ºC stored up to 36-h.

Conclusion:

Serum gel or non-gel tubes might be used interchangeably for 11 analytes chilled or at 24 ºC, whereas some restrictions must be applied for glucose, AST, BUN, HDL, and uric acid.  相似文献   

18.

Introduction

Hemolysis can occur during sample collection, handling and transport. It is more frequent when the non-laboratory staff performs sampling. The aim of this study was to assess nurses’ knowledge on the causes of hemolysis and consequential impact on the laboratory tests results. Additionally, the differences in knowledge, related to work experience, professional degree and previous education about hemolysis were explored.

Materials and methods

An anonymus survey, containing 11 questions on demographics, causes of hemolysis, its impact on biochemical parameters and nurses’ attitude towards additional education in preanalytics, was conducted in four Croatian hospitals. The answers were compared by Chi-squared and Fischer exact test.

Results

In total, 562 survey results were collected. Majority of nurses declared familiarity with the term “hemolysis” (99.6%). There were 77% of correct answers regarding questions about the causes of hemolysis, but only 50% when it comes to questions about interference in biochemical tests. The percentage of correct answers about causes was significantly lower (P = 0.029) among more experienced nurses, and higher (P = 0.027) in those with higher professional degree, while influence of previous education was not significant. Also, higher percentage of correct answers about interferences was encountered in nurses with longer work experience (P = 0.039). More than 70% of nurses declared that additional education about preanalytical factors would be beneficial.

Conclusion

Croatian nurses are familiar with the definition of hemolysis, but a lack of knowledge about causes and influence on laboratory test results is evident. Nurses are eager to improve their knowledge in this field of preanalytical phase.Key words: hemolysis, nurses, survey, preanalytical phase  相似文献   

19.

Introduction

The study of cardiac response to strenuous and continuous exercise is crucial to understanding the physiology of endurance. N-terminal proB-type natriuretic peptide (NT-proBNP) is a potential marker for monitoring myocardial wall stress, and troponins (TnT and TnI) are widely used in the diagnosis of cardiac ischemia and infarction. Strenuous exercise may generate transitory ischemia, myocardial stress, and diastolic left ventricular dysfunction, inducing the increased production of both these biomarkers. We measured changes in NT-proBNP and TnT in elite cyclists during a 3-week stage race, a model of strenuous exercise.

Materials and methods:

The study population was 9 professional cyclists participating in the 2011 Giro d’Italia. Pre-analytical and analytical phases scrupulously followed official recommendations. Anthropometric data, net energy expenditure and cardiac indexes (rate, diastolic and systolic blood pressure) were recorded. Blood samples were drawn pre-race (day −1) and at days 12 and 22; NT-proBNP and highly sensitive-troponin (Hs-TnT) concentrations were assayed and corrected for plasma volume changes.

Results:

Body-mass index decreased and energy expenditure increased by 52% during the race. NT-proBNP concentrations increased [day −1: 23.52 ng/L (9.67–34.33); day 12: 63.46 ng/L (22.15–93.31); P = 0.039; day 22: 89.26 ng/L (34.66–129.78) vs. day −1; P < 0.001] and correlated with heart rate (r = −0.51; P = 0.006), systolic pressure (r = 0.39; P = 0.046) and energy expenditure (r = 0.70; P < 0.001). TnT concentrations did not vary, but a widened TnT amplitude distribution was observed.

Conclusions:

Increases in NT-proBNP correlated with higher energy expenditure over a 3-week cycling stage race, possibly indicating myocardial stress.  相似文献   

20.

Introduction:

Today, the pneumatic tube transport system (PTS) is used frequently because of its advantages related to timing and speed. However, the impact of various types of PTS on blood components is unknown. The aim of this study was to examine the influence of PTS on the quality of routine blood cell counts, erythrocyte sedimentation, and certain blood coagulation tests.

Materials and methods:

Paired blood samples were obtained from each of 45 human volunteers and evaluated by blood cell count, erythrocyte sedimentation, and several coagulation tests, including prothrombin time (PT) and activated partial thromboplastin time (aPTT). Blood samples were divided into 2 groups: Samples from group 1 were transported to the laboratory via the PTS, and samples from group 2 were transported to the laboratory manually. Both groups were evaluated immediately by the tests listed above.

Results:

The blood sample test results from groups 1 and 2 were evaluated and compared. No statistically significant differences were observed (P = 0.069–0.977).

Conclusion:

The PTS yielded no observable effects on blood cell counts, erythrocyte sedimentation, or PT and aPTT test results. We concluded that the PTS can be used to transport blood samples and yield reliable results for blood cell counts, erythrocyte sedimentation, and several coagulation tests.  相似文献   