INFRARED MICROSPECTROSCOPY FOR THE ANALYSIS OF OLD PAINTING MATERIALS
Abstract:Abstract

The Perkin-Elmer infrared microspectrometer consists of a basic, double-beam spectrometer M 13, on which a microscopical system model 85 with reflecting optics can be mounted. The microspectrometer can be used only for single-beam measurements. Pressed in a KBr-pellet, with a radius of 0·5 mm, the amount of sample of a pure material can be decreased to 2 μg. A sample with different paint-layers can be embedded in a wax with little infrared absorption and can be cut with the help of a microtome in slices with a thickness of 20 μ. Every paint-layer can be measured separately by means of an adjustable diaphragm in the microscope. A sample area of 37·5 ? 500 μ is sufficient to obtain an identifiable spectrum.