首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Technical Reliability Assessment of the Actigraph GT1M Accelerometer
Authors:Pedro Silva  Jorge Mota  Dale Esliger  Gregory Welk
Institution:1. Research Centre in Physical Activity, Health and Leisure , University of Porto , Porto, Portugal perrinha@gmail.com;3. Research Centre in Physical Activity, Health and Leisure , University of Porto , Porto, Portugal;4. School of Sport and Health Sciences, University of Exeter , Exeter, UK;5. Kinesiology Department , Iowa State University , Ames, Iowa
Abstract:The purpose of this study was to determine the reliability of the Actigraph GT1M (Pensacola, FL, USA) accelerometer activity count and step functions. Fifty GT1M accelerometers were initialized to collect simultaneous acceleration counts and steps data using 15-sec epochs. All reliability testing was completed using a mechanical shaker plate to perform six different test conditions in Experiment 1 and 18 test conditions in Experiment 2. The overall intra- and inter-instrument reliability of the GT1M was CVintra = 2.9% and CVinter = 3.5% for counts and CVintra = 1.1% and CVinter = 1.2% for steps. No batch effects were evident in the 50 GT1Ms. The Actigraph GT1M accelerometer demonstrated good reliability for measuring both counts and steps. However, the ability of the GT1M to consistently detect acceleration at a given acceleration and frequency condition varied widely. Future studies clarifying the filtering limitations and the threshold necessary to detect the occurrence of movement are warranted.
Keywords:frequency  pedometer  filter  acceleration