Validation of the use of foreign gas rebreathing method for non-invasive determination of cardiac output in heart disease patients
Authors:Dong?Liang  Email author" target="_blank">Wang?Jian-an?Email author  Jiang?Chen-yang
Institution:(1) Department of Cardiology, Sir Run Run Shaw Hospital, School of Medicine, Zhejiang University, Hangzhou, 310016, China
Abstract:Objective: To compare a new device (Innocor) for non-invasive measurement of cardiac output (CO) by foreign gas rebreathing method with conventional techniques used in the measurements of cardiac function. Methods: Cardiac outputs measured by Innocor (CORB) were compared with CO obtained by echocardiography (COEC), Swan-Ganz thermodilution (COTD),and left ventricle radiography (COLVR) in 34 patients subjected to cardiac catheterization. Values obtained from the four methods were analyzed by linear regression and paired values were compared by the method of Bland and Altman in SPSS. Results: There was strong positive correlation (r=0.94) between Innocor cardiac output values and the corresponding values obtained by thermodilution and between COEC and COLVR values. Thermodilution appears to overestimate cardiac output when compared to the values obtained with Innocor by (0.66±0.22) L/min (P<0.0001). There was no correlation between data obtained by Innocor and the corresponding COEC and COLVR values. Conclusion: Innocor CORB is an easy, safe and well established method for non-invasive measurement of cardiac output with good prospects for clinical application in heart disease patients.
Keywords:Indirect Fick (foreign gas rebreathing)  Swan-Ganz  Cardiac output  Left ventricle radiography  Echocardiography
本文献已被 CNKI 万方数据 PubMed SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号