首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Lack of precision in HbA1c values on variant II in cases of hemoglobin Q India
Authors:P K Nigam  S Sharma  R Sareen  V Paul  A Lal
Institution:(1) Dr Lal PathLabs Pvt Ltd, ‘ESKAY HOUSE’, 54, Hanuman Road, 110001 New Delhi, India
Abstract:This study was carried out to check the precision of HbA1c values on Bio-Rad Variant II in cases of a rare hemoglobin variant Q India. The study was carried out over a three month period on samples collected for HbA1c estimation. Seven out of eleven patients showed variable results of HbA1c with a very high and unacceptable intraday mean coefficient of variation (CV) of 9.93%. We conclude, that the results of HbA1c on Variant II can not be reported without adversely affecting HbA1c as a marker of long-term glycemic control in patients who have hemoglobin Q India. The HbA1c value of these patients needs to be assessed by a different instrument/method or the glycemic control be monitored by an alternate test like serum Fructosamine.
Keywords:HbA1c            precision  variant hemoglobin  diabetes mellitus
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏