首页 | 本学科首页   官方微博 | 高级检索  
     检索      


An integrated microfluidic cell array for apoptosis and proliferation analysis induction of breast cancer cells
Authors:Song Huixue  Chen Tan  Zhang Baoyue  Ma Yifan  Wang Zhanhui
Institution:Institute of Biomedical and Health Engineering, Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences, Shenzhen 518055, China and Key Laboratory of Biomedical Informatics and Health Engineering, Chinese Academy of Sciences, 1068 Xueyuan Avenue, Shenzhen 518055, China.
Abstract:In vitro sensitivity testing of tumor cells could rationalize and improve the choice of chemotherapy and hormone therapy. In this report, a microfluidic device made from poly(dimethylsiloxane) and glass was developed for an assay of drug induced cytotoxicity. We evaluated the apoptotic and proliferation-inhibitory effects of anticancer drugs mitomycin C (MMC) and tamoxifen (TAM) using MCF-7 breast cancer cells. MMC and TAM both induced apoptosis and inhibited proliferation of MCF-7 cells in a concentration-dependent manner. MMC caused the expression of antiapoptotic protein Bcl-2 a dose-dependent reduction in MCF-7 cells. The expression of Bcl-2 did not change significantly in MCF-7 cells treated by TAM. The results in the microfluidic device were correlated well with the data obtained from the parallel experiments carried out in the conventional culture plates. The developed microfluidic device could be a potential useful tool for high content screening and high throughput screening research.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏